FDA Adverse Event
Injury
Summary report: N
840 VENTILATOR
MDR report key: 4092781
·
Received August 19, 2014
Report
- Report Number
- 8020893-2014-02137
- Event Type
- Injury
- Date Received
- August 19, 2014
- Date of Event
- August 16, 2014
- Report Date
- August 18, 2014
- Manufacturer
- COVIDIEN
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) EVALUATED THE DEVICE, AND VERIFIED THE REPORTED MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB), THE BREATH DELIVERY UNIT (BDU) PCB, AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE UNIT PASSED ALL TESTS AND CALIBRATIONS, AND IT WAS OBSERVED TO BE OPERATING WITHIN MANUFACTURING SPECIFICATIONS. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED BY A CUSTOMER IN (B)(6) THAT, DURING A PATIENT USE, THE 840 VENTILATOR GRAPHIC USER INTERFACE (GUI) DISPLAYED LOSS OF COMMUNICATION ERROR CODES. THE PATIENT WAS TRANSFERRED TO ANOTHER VENTILATOR. THERE WAS NO REPORTED PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 497008 | 840 VENTILATOR | CONTINUOUS VENTILATOR | CBK | COVIDIEN | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |