FDA Adverse Event Malfunction Summary report: N

TRILOGY 200

MDR report key: 4092770 · Received August 20, 2014

Report

Report Number
2518422-2014-01384
Event Type
Malfunction
Date Received
August 20, 2014
Date of Event
July 25, 2014
Report Date
July 25, 2014
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K093416
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CTR, THE DEVICE FAILED STEPS DURING TESTING. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUES.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MFR FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
500583 TRILOGY 200 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1032804