FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 4092758 · Received August 20, 2014

Report

Report Number
2518422-2014-01402
Event Type
Malfunction
Date Received
August 20, 2014
Date of Event
July 23, 2014
Report Date
July 23, 2014
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CTR, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S POWER MANAGEMENT BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

THE MFR RECEIVED INFO ALLEGING A "SERVICE REQUIRED" ALARM CONDITION. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
500704 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1054655

Patients

Seq Age Sex Outcome Treatment
1