Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

FURTHER INVESTIGATION EFFORTS FOR CUSTOMER REPORTS OF THE CUFF NOT DEFLATING WERE COMPLETED ON (B)(4) 2014. INVESTIGATION RESULTS REVEALED THE FOLLOWING: INVESTIGATION OF DEFLATION ISSUES ASSOCIATED WITH THE BROCHOCATH PRODUCTS HAS DEMONSTRATED THAT THE POLYURETHANE MATERIAL OF THE TRACHEAL CUFF MAY COLLAPSE IN A NON-SYMMETRICAL MANNER DURING DEFLATION. THIS MAY CREATE A SEAL AROUND THE PERIMETER OF THE NOTCHES IN THE CATHETER, WHICH MAY BLOCK THE PATH FOR AIR EXTRACTION DURING CUFF DEFLATION. ADDITIONALLY, THE INVESTIGATION HAS SHOWN THAT THE DIFFICULTY ASSOCIATED WITH CUFF DEFLATION IS MOST LIKELY TO OCCUR WHEN THE TUBE IS BENT NEAR THE INTERFACE BETWEEN THE NOTCH OPENING AND THE CUFF WALL, AND WHEN THE NOTCHED OPENINGS ARE ORIENTED TO THE OUTSIDE OF THE CURVATURE. INFORMATION HAS BEEN ADDED TO THE DATABASE, AND TRENDS WILL CONTINUE TO BE MONITORED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY A CUSTOMER IN (B)(6) THAT, DURING PRIOR TO USE TESTING, THE ENDOBRONCHIAL TUBE CUFF WOULD NOT DEFLATE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
496940 MALLINCKRODT ENDOBRONCHIAL TUBE BTS COVIDIEN 210401452