FDA Adverse Event Malfunction Summary report: N

VERSACARE BED

MDR report key: 4092613 · Received September 8, 2014

Report

Report Number
1824206-2014-02210
Event Type
Malfunction
Date Received
September 8, 2014
Date of Event
August 11, 2014
Report Date
August 11, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE BED'S SCALE HAD BEEN READING INACCURATELY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNK IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE FOOT RIGHT LOAD BEAM TO RESOLVE THE ISSUE. THE TECHNICIAN THOROUGHLY TESTED THE BED EXIT SYS AND IT FUNCTIONED AS DESIGNED. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE SCALE WAS READING INACCURATE WEIGHTS. THE BED WAS LOCATED IN ROOM 408 AT THE ACCOUNT. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYS AS COMPLAINT #180727.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
547546 VERSACARE BED A/C POWERED ADJUSTABLE HOSP BED FNL HILL-ROM, INC. 3200

Patients

Seq Age Sex Outcome Treatment
1