FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 4092609 · Received September 8, 2014

Report

Report Number
1824206-2014-02219
Event Type
Malfunction
Date Received
September 8, 2014
Date of Event
August 11, 2014
Report Date
August 11, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE SIDE RAIL SLIDE ASSEMBLY WAS DAMAGED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2012-2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE SIDE RAIL SLIDE ASSEMBLY TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE HEAD RIGHT SIDE RAIL WOULD NOT LATCH. THE BED WAS LOCATED IN BED STORAGE AT THE ACCOUNT. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
547607 ADVANTA BED A/C POWERED ADJUSTABLE HOSP BED FNL HILL-ROM, INC. 1600

Patients

Seq Age Sex Outcome Treatment
1