ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-26354
- Event Type
- Malfunction
- Date Received
- September 16, 2014
- Report Date
- September 3, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON 10/31/2014 WITH THE FOLLOWING FINDINGS: THE REPORTED NO-POWER ISSUE COULD NOT BE ADEQUATELY INVESTIGATED DUE TO A BLANK-DISPLAY ISSUE, AND THUS NO CONCLUSIONS COULD BE DETERMINED IN THIS REGARD. THERE WERE NO POWER-REBOOT EVENTS, WHICH WOULD BE PRESENT IN THE CASE OF A NO-POWER ISSUE, OBSERVED IN THE BLACK BOX DATA. HOWEVER, A CS-054 CALL SERVICE ALARM, WHICH IS OFTEN MISTAKEN FOR A POWER FAILURE, WAS OBSERVED IN THE BLACK BOX DATA FROM THE ALERT DATE. MOISTURE WAS PRESENT BEHIND THE DISPLAY LENS. THE PUMP CASE WAS CRACKED NEAR THE RIGHT SIDE OF THE DISPLAY LENS. THE PUMP FAILED A LEAK TEST DUE TO THE PUMP CASE DAMAGE; THIS DEVICE FAILURE CAUSED THE MOISTURE INGRESS EVENT. THE PUMP POWERED ON, AS CONFIRMED BY THE PRESENCE OF THE AUDITORY AND VIBRATORY CUES; HOWEVER, THE DISPLAY SCREEN WAS BLANK DUE TO INTERNAL MOISTURE DAMAGE. THE PUMP CASE WAS REMOVED, AND MOISTURE DAMAGE WAS FOUND ON INTERNAL COMPONENTS. THE RELATION OF THE AFOREMENTIONED DEVICE FAILURES TO THE REPORTED NO-POWER ISSUE CAN NOT BE DETERMINED, AS THE REPORTED NO-POWER ISSUE WOULD HAVE TO BE DUPLICATED TO DETERMINE SUCH. UNRELATED TO THE REPORTED NO-POWER ISSUE, THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED BELOW THE GRIP PAD.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. IT WAS REPORTED THAT MOISTURE INGRESS WAS EVIDENT BEHIND THE DISPLAY LENS AND THE PUMP WAS EXPERIENCING A NO-POWER ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 573654 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |