FDA Adverse Event Malfunction Summary report: N

HUDSON CONCHA NEPTUNE

MDR report key: 4092538 · Received August 19, 2014

Report

Report Number
3003898360-2014-00522
Event Type
Malfunction
Date Received
August 19, 2014
Date of Event
July 29, 2014
Report Date
July 29, 2014
Manufacturer
TELEFLEX
Product Code
BTT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A VISUAL, FUNCTIONAL AND DIMENSIONAL INSPECTION OF THE PRODUCT INVOLVED IN THE COMPLAINT COULD NOT BE CONDUCTED SINCE THE PRODUCT WAS NOT RETURNED. PER DHR )DEVICE HISTORY RECORD) THE PRODUCT CONCHA NEPTUNE, (B)(4) WAS MANUFACTURED ON 06/20/2008. THE DHR INVESTIGATION DID NOT SHOW ISSUES RELATED TO COMPLAINT. A DOCUMENT ASSESSMENT (FMEA) WAS CONDUCTED AND NO CHANGES WERE REQUIRED. CORRECTIVE ACTIONS CAN NOT BE ESTABLISHED SINCE IT IS NECESSARY TO RECEIVE THE PHYSICAL SAMPLE TO PERFORM A PROPER INVESTIGATION AND CONFIRM THE DEFECT. AT THIS TIME SINCE THE DEVICE PRODUCT IS NOT AVAILABLE IT IS NOT POSSIBLE TO DETERMINE THE SOURCE OF THE DEFECT REPORTED. THE COMPLAINT CAN NOT BE CONFIRMED SINCE THE SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION, THEREFORE, IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE FOR THE DEFECT REPORTED AND A CORRECTIVE ACTION FOR IT. IF THE DEVICE SAMPLE BECOMES AVAILABLE AT A LATER DATE, THIS COMPLAINT WILL BE UPDATED ACCORDINGLY. TELEFLEX WILL CONTINUE TO MONITOR AND TREND RELATING COMPLAINTS.

Description of Event or Problem · 1

THE EVENT IS REPORTED AS: THE UNIT DISPLAYS A RED WRENCH (INTERNAL FAILURE) AT THE TIME OF SET-UP. NO PT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
500085 HUDSON CONCHA NEPTUNE HEATED HUMIDIFICATION SYSTEM BTT TELEFLEX

Patients

Seq Age Sex Outcome Treatment
1