FDA Adverse Event Malfunction Summary report: N

GREENLIGHT MOXY FIBER OPTIC

MDR report key: 4092528 · Received September 16, 2014

Report

Report Number
2937094-2014-00836
Event Type
Malfunction
Date Received
September 16, 2014
Date of Event
August 8, 2014
Report Date
August 11, 2014
Manufacturer
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
Product Code
GEX
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER 0010-2400-416A-(B)(4). THE FIBER SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE DISTAL TO THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE METAL CAP EXHIBITS MODERATE CHAR; THE GLASS CAP EXHIBITS SEVERE DEVITRIFICATION AT THE OUTPUT WINDOW; THE METAL CAP ADHESION LOCATION EXHIBITS BURNT GLUE; THE OUTER FLOW TUBING EXHIBITS MILD CONTAMINATION, LIKELY BIOLOGIC. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER. (B)(4) REFERS TO FORWARD-FIRING OF THE SIDE-FIRING SURGICAL FIBER; A CODE REQUEST HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROSTATE PROCEDURE, "FIBER START GOING ON AND OFF OVER AND OVER DURING THE PROCEDURE UNTIL THE SYSTEM REQUEST A NEW FIBER", AT 89,934 JOULES AND 13:23 MINUTES OF USAGE.THE FIBER WAS EXCHANGED AND THE CASE COMPLETED. PATIENT OUTCOME: ¿OK¿ REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
573056 GREENLIGHT MOXY FIBER OPTIC POWERED LASER SURGICAL INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) 0010-2400 416A

Patients

Seq Age Sex Outcome Treatment
1