GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2014-00836
- Event Type
- Malfunction
- Date Received
- September 16, 2014
- Date of Event
- August 8, 2014
- Report Date
- August 11, 2014
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER 0010-2400-416A-(B)(4). THE FIBER SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE DISTAL TO THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE METAL CAP EXHIBITS MODERATE CHAR; THE GLASS CAP EXHIBITS SEVERE DEVITRIFICATION AT THE OUTPUT WINDOW; THE METAL CAP ADHESION LOCATION EXHIBITS BURNT GLUE; THE OUTER FLOW TUBING EXHIBITS MILD CONTAMINATION, LIKELY BIOLOGIC. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER. (B)(4) REFERS TO FORWARD-FIRING OF THE SIDE-FIRING SURGICAL FIBER; A CODE REQUEST HAS BEEN SUBMITTED.
IT WAS REPORTED THAT DURING A PROSTATE PROCEDURE, "FIBER START GOING ON AND OFF OVER AND OVER DURING THE PROCEDURE UNTIL THE SYSTEM REQUEST A NEW FIBER", AT 89,934 JOULES AND 13:23 MINUTES OF USAGE.THE FIBER WAS EXCHANGED AND THE CASE COMPLETED. PATIENT OUTCOME: ¿OK¿ REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 573056 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 416A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |