FDA Adverse Event Injury Summary report: N

INFINIA HAWKEYE

MDR report key: 4092453 · Received September 4, 2014

Report

Report Number
9613299-2014-00004
Event Type
Injury
Date Received
September 4, 2014
Date of Event
August 6, 2014
Report Date
August 11, 2014
Manufacturer
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Product Code
KPS
PMA / PMN Number
K052434
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

GE HEALTHCARE'S INVESTIGATION IS ONGOING. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PT WAS POSITIONED WITH HER ARMS PROTRUDING FROM THE SIDES OF THE TABLE, AND A CHAIR WAS PLACED WITHIN THE DETECTOR'S OPERATION AREA. AS PART OF THE AUTOMATIC UNLOAD SEQUENCE INITIATED BY THE OPERATOR, WHILE THE DETECTORS WERE ROTATING, THE PT'S HAND WAS TRAPPED BETWEEN THE SIDE OF THE DETECTOR AND THE CHAIR. AS A RESULT THE PT'S HAND WAS "SQUEEZED," AND THE PT SUFFERED OF FIVE BROKEN BONES ON HER HAND REQUIRING A CAST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
541637 INFINIA HAWKEYE TOMOGRAPHY, COMPUTED, EMISSION KPS GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING

Patients

Seq Age Sex Outcome Treatment
1 55 YR Other| R