FDA Adverse Event Malfunction Summary report: N

STERLING?

MDR report key: 4092434 · Received September 16, 2014

Report

Report Number
2134265-2014-05454
Event Type
Malfunction
Date Received
September 16, 2014
Date of Event
August 20, 2014
Report Date
August 20, 2014
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
DQY
PMA / PMN Number
K053118
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). DEVICE EVALUATED BY MFR: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED EXTERNAL ILIAC ARTERY. THE 6.0MMX60MMX135CM (4F) STERLING¿ BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY SEGMENT OF THE DEVICE RETAINED INSIDE THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
572887 STERLING? CATHETER, PERCUTANEOUS DQY BOSTON SCIENTIFIC - MAPLE GROVE H74939031606010 16371531

Patients

Seq Age Sex Outcome Treatment
1 GUIDE WIRE: CHEVALIER