FDA Adverse Event Malfunction Summary report: N

THUNDERBEAT 5MM, 35CM FRONT ACTUATED GRIP

MDR report key: 4092393 · Received September 5, 2014

Report

Report Number
8010047-2014-00519
Event Type
Malfunction
Date Received
September 5, 2014
Date of Event
August 11, 2014
Report Date
August 12, 2014
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
LFL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) FOR EVALUATION. THE EVALUATION CONFIRMED THAT THE PTFE PAD WAS PARTIALLY PEELED FROM THE JAW. THERE WERE CONTACT MARKS ON THE SURFACE OF THE PROBE AND THE JAW. THE MANUFACTURING RECORD WAS REVIEWED WITH NO IRREGULARITIES. BASED ON THE SIMILAR CASES WITH THE SAME MODEL, IT IS KNOWN THAT BY CONTINUOUSLY ACTIVATING OUTPUT WITHOUT GRASPING ANYTHING IN THE GRASPING SECTION (INCLUDING AFTER THE TISSUE SEPARATED), THE PTFE PAD IS SEVERELY WORN, AND THE DISTAL END OF THE PAD IS SEPARATED FROM THE GRASPING SECTION. CONSIDERING THE EVALUATION RESULT OF THE SUBJECT DEVICE, OMSC CONCLUDED THAT THE REPORTED EVENT OCCURRED SINCE THE USER CONTINUED ACTIVATING OUTPUT FOR AN EXTENDED TIME AFTER THE TISSUE ALREADY CUT. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.

Description of Event or Problem · 1

THE SUBJECT DEVICE WAS USED DURING A LAPAROSCOPIC COLECTOMY. WHEN THE SURGEON OUTPUT IN SEAL AND CUT MODE, HE FELT THAT SOMETHING WAS WRONG AND IT WAS FOUND THAT THE PTFE PAD OF THE DEVICE WAS SEPARATED. THE PROCEDURE WAS COMPLETED WITH ANOTHER THUNDERBEAT DEVICE. THERE WAS NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
545987 THUNDERBEAT 5MM, 35CM FRONT ACTUATED GRIP THUNDERBEAT HANDPIECE LFL OLYMPUS MEDICAL SYSTEMS CORPORATION TB-0535FC K4131

Patients

Seq Age Sex Outcome Treatment
1