FDA Adverse Event Injury Summary report: N

840 VENTILATOR

MDR report key: 4092330 · Received August 19, 2014

Report

Report Number
8020893-2014-02035
Event Type
Injury
Date Received
August 19, 2014
Date of Event
August 1, 2014
Report Date
August 1, 2014
Manufacturer
COVIDIEN
Product Code
CBK
PMA / PMN Number
K970460
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
RESPIRATORY THERAPIST

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) EVALUATED THE DEVICE, AND WAS UNABLE TO DUPLICATE THE REPORTED MALFUNCTION. THERE WERE NO ERROR CODES IN THE MEMORY LOG, AND NO PARTS WERE REPLACED. THE UNIT PASSED ALL TESTS AND CALIBRATIONS, AND IT WAS OBSERVED TO BE OPERATING WITHIN MANUFACTURING SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY A CUSTOMER IN THE USA THAT, DURING PATIENT USE, THE 840 VENTILATOR GRAPHIC USER INTERFACE (GUI) DISPLAY WAS BLANK. THE PATIENT WAS TRANSFERRED TO ANOTHER VENTILATOR. THERE WAS NO REPORTED PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
500269 840 VENTILATOR CONTINUOUS, VENTILATOR CBK COVIDIEN 840

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention