CHARGING SYSTEM
Report
- Report Number
- 1627487-2014-02593
- Event Type
- Injury
- Date Received
- August 28, 2014
- Date of Event
- August 8, 2014
- Report Date
- August 8, 2014
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 3 OF 3. REFERENCE MFR REPORT: 1627487-2014-02591, 16267487-2014-02592. THE MANUFACTURER'S PATIENT DATABASE INDICATES THE PATIENT HAS TWO 3701 MODEL CHARGERS. IT IS UNDETERMINED WHICH CHARGER CAUSED THE HEATING ISSUE. THEREFORE, ALL POSSIBLE DEVICES ARE BEING REPORTED. IT WAS REPORTED THE PATIENT IS EXPERIENCING DIFFICULTY COMMUNICATING WITH HIS SCS IPG AND THE CHARGER. ALSO, THE PATIENT STATED HE EXPERIENCED POCKET HEATING WHEN CHARGING HIS IPG. A REPLACEMENT CHARGING SYSTEM WAS SENT TO THE PATIENT TO ADDRESS THE ISSUE. ST. JUDE MEDICAL, NEUROMODULATION DIVISION, HAS SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 526437 | CHARGING SYSTEM | SCS CHARGING SYSTEM | GZB | ST. JUDE MEDICAL - NEUROMODULATION | 3701 | 4275829 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | Other |