FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 4091761 · Received September 16, 2014

Report

Report Number
2024168-2014-05948
Event Type
Injury
Date Received
September 16, 2014
Date of Event
August 20, 2014
Report Date
August 25, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE OTHER PERCLOSE PROGLIDE DEVICE IS BEING FILED UNDER SEPARATE MEDWATCH MFR NUMBER. EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED UNFORMED/ UNRAVELED KNOT COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS THE PLUNGER, LINK, NON-RAIL SUTURE CONTAINING THE KNOT, NEEDLES AND CUFFS WERE NOT RETURNED AND 12 INCHES OF RAIL SUTURE WAS RETURNED OUTSIDE THE DEVICE AND ITS END WAS CUT. THE REPORTED UNFORMED KNOT COULD NOT BE CONFIRMED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES FOR THIS LOT. A QUERY OF THE ELECTRONIC COMPLAINT HANDLING DATABASE REVEALED NO SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE WITH A 6FR SHEATH, AFTER A CORONARY INTERVENTIONAL PROCEDURE. REPORTEDLY, A KNOT DID NOT FORM. A SECOND PROGLIDE DEVICE WAS USED WITH THE SAME RESULTS. MANUAL ARTERIAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
572765 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 40617K1

Patients

Seq Age Sex Outcome Treatment
1 51 YR Required Intervention SHEATH: 6FR