MINI AIR DRILL
Report
- Report Number
- 8030965-2014-01343
- Event Type
- Malfunction
- Date Received
- September 16, 2014
- Report Date
- November 22, 2012
- Manufacturer
- SYNTHES GMBH
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL NARRATIVE: SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. PLACEHOLDER.
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE BEARING, CASING, AND LEAVES ARE WORN OUT. THE STARTING LEVER IS LOOSE AND THE TUBE IS LEAKING. THIS WAS ATTRIBUTED TO NORMAL WEAR. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER ON (B)(4) 2012. PLACEHOLDER.
CUSTOMER SENT THE DEVICE IN FOR REPAIR. THIS IS REPORT 1 OF 1 FOR COMPLAINT#(B)(4).
CUSTOMER SENT THE DEVICE IN FOR REPAIR. TECHNICIAN FOUND THAT THE STARTING LEEVER WAS LOOSE, THE CASING AND LEAVES WORN OUT, THE TUBE WAS LEAKING AND THE CONNECTION RING WAS PARTLY CHIPPED. THIS IS REPORT 1 OF 1 FOR COMPLAINT #(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 571661 | MINI AIR DRILL | HWE | SYNTHES GMBH | 8140 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |