PIN REV DOME DRILL 25MM
Report
- Report Number
- 1818910-2014-28127
- Event Type
- Malfunction
- Date Received
- September 16, 2014
- Date of Event
- September 8, 2014
- Report Date
- September 8, 2014
- Manufacturer
- DEPUY ORTHOPAEDICS, INC. 1818910
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A COMPLAINT DATABASE SEARCH HAS IDENTIFIED A TREND FOR THIS FAILURE WITHIN THE 2274 QUICKSET DRILL FAMILY. PREVIOUS INVESTIGATIONS HAVE DETERMINED THAT USE ERROR IS THE LIKELY ROOT CAUSE. AS A RESULT OF TRENDING HEALTH HAZARD EVALUATION, DVA-106292-HHE WAS CONDUCTED. THE INVESTIGATION DID NOT ESTABLISH THE NEED FOR CORRECTIVE ACTION. NO EVIDENCE WAS FOUND OF PRODUCT OR DESIGN ERROR AS A CONTRIBUTING FACTOR. SUBSEQUENT COMPLAINTS WILL BE MONITORED UNDER SEP 419 POST MARKET SURVEILLANCE. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY..
INSTRUMENT BROKE DURING SURGERY. NO REP WAS IN ATTENDANCE WHEN THIS HAPPENED AND HOSPITAL HAVE DISCARDED THE INSTRUMENT. NO DELAY TO SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 571351 | PIN REV DOME DRILL 25MM | HIP INSTRUMENT/TRIAL | LXH | DEPUY ORTHOPAEDICS, INC. 1818910 | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |