Description of Event or Problem · 1

PER INDEPENDENT REPAIR CENTER STATEMENT, THE UNIT HAS LOW O2 OR YELLOW LIGHT. THE KEY FAILURE IS THE PILOT VALVE FOR THE MANIFOLD IS CONTAMINATED. ADDITIONAL MALFUNCTIONS ARE THE ON/OFF SWITCH FOR THE CONTROL PANEL HAS A SHORT CIRCUIT, THE INTAKE FILTER FOR THE SOUND BOX IS DIRTY/STAINED, AND THE PILOT VALVE O-RING AND TIE STRAPS FOR THE SIEVE BEDS ARE DEFECTIVE.