OT VERIO IQ METER
Report
- Report Number
- 2939301-2014-24218
- Event Type
- Malfunction
- Date Received
- September 16, 2014
- Report Date
- September 8, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 ¿ (11/12/2014). THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON 10/7/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 10/28/2014 WITH THE FOLLOWING FINDINGS:THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE TEST STRIPS WERE FOUND TO HAVE RESULTS GREATER THAN +50% OF THE CONTROL SOLUTION WHEN TESTED WITH CONTROL SOLUTION. IN ADDITION, A SECONDARY ISSUE WAS NOTED WHEN THE TEST STRIPS WERE FOUND TO HAVE SCORE ACROSS THE BACK OF THE ELECTRODES. ADDITIONAL TESTS WERE PERFORMED ON THE TEST STRIP VIAL AND THE DESICCANT TO CHECK THE STRUCTURAL INTEGRITY AND FOR POSSIBLE MOISTURE ISSUE. THE STRUCTURAL INTEGRITY OF THE TEST STRIP VIAL WAS FOUND TO BE INTACT DURING A GROSS LEAK TEST. THE DESICCANT WITHIN THE SUBJECT VIAL WAS FOUND TO BE SATURATED BY MOISTURE AND FAILED TGA TESTING. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN (B)(4) ALLEGING THAT THE SUBJECT METER (ONETOUCH VERIO IQ) READ INACCURATELY HIGH COMPARED TO ANOTHER DEVICE. THE REPORTER CLAIMED OBTAINING A BLOOD GLUCOSE READING OF "30MMOL/L" WITH THE SUBJECT METER AND "6.7MMOL/L" ON ANOTHER DEVICE, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN¿S CRITERIA FOR ACCURACY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 571316 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3615590 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |