ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-26334
- Event Type
- Malfunction
- Date Received
- September 15, 2014
- Report Date
- September 2, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/28/2014 WITH THE FOLLOWING FINDINGS: CS 064-0001 ALARM VERIFIED IN BLACK BOX AND HISTORY. NO ALARMS DURING INVESTIGATION. OPENED PUMP. NO DEFECTS FOUND. BOTH CAPS RETURNED AND USED TO COMPLETE INVESTIGATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 064) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 569027 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |