ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-26322
- Event Type
- Malfunction
- Date Received
- September 15, 2014
- Report Date
- September 2, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1 - PRODUCT ANALYSIS: THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 04/08/2015 WITH THE FOLLOWING FINDINGS: THE COMPLAINT WAS UNABLE TO BE DUPLICATED OR CONFIRMED WITH INVESTIGATION. REVIEW OF THE PUMP¿S BLACK BOX FOUND NO EVIDENCE OF RELATED ISSUES; DATA RELEVANT TO THE ALLEGED ISSUE WAS OVERWRITTEN DUE TO CONTINUED PUMP USE. THE PUMP SUCCESSFULLY COMPLETED A PRIME SEQUENCE AND 24-HOUR EXERCISE TEST WITHOUT ISSUE OR ALARM. THE PUMP¿S FORCE SENSOR CALIBRATION WAS WITHIN SPECIFICATION. NO DAMAGE OR DEFECT WAS FOUND TO THE PUMP¿S INTERNAL COMPONENTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A LOSS OF PRIME ISSUE. NO ADDITIONAL INFORMATION WAS PROVIDED AND TROUBLESHOOTING WAS UNABLE TO BE COMPLETED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 568985 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 YR |