FDA Adverse Event Malfunction Summary report: N

EMAX 2 PLUS MOTOR

MDR report key: 4090764 · Received September 15, 2014

Report

Report Number
1045834-2014-12831
Event Type
Malfunction
Date Received
September 15, 2014
Date of Event
September 2, 2014
Report Date
September 3, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK080802
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE CORD HAD A SLIT IN IT WHICH IS INDICATIVE OF ALLOWING THE CORD TO COME INTO CONTACT WITH A SHARP OBJECT. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO MISUSE, ABUSE AND OR ERROR. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS OBSERVED THAT THE BLACK CORD ON THE MOTOR DEVICE HAD A CUT AND WIRES WERE EXPOSED. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE. A SPARE IDENTICAL DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
568745 EMAX 2 PLUS MOTOR MOTOR, DRILL, ELECTRIC - HANDPIECE HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1