11CM ANGLE ATTACHMENT
Report
- Report Number
- 1045834-2014-12846
- Event Type
- Malfunction
- Date Received
- September 15, 2014
- Report Date
- September 2, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(6). AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE HAD DAMAGED BEARINGS. IT WAS DETERMINED THAT THIS WOULD CAUSE HEAT. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL USE AND SERVICING OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
THIS IS REPORT 2 OF 2 FOR THE SAME EVENT. IT WAS REPORTED FROM GERMANY THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE MOTOR DEVICE WAS HEATING UP WHEN IN USE WITH AN ATTACHMENT DEVICE. THE SURGEON COULD NOT DETERMINE IF THE HEATING WAS CAUSED FROM THE MOTOR DEVICE DIRECTLY OR THE ATTACHMENT DEVICE. THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS SPARE DEVICES WERE AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 569844 | 11CM ANGLE ATTACHMENT | MOTOR, DRILL, ELECTRIC - ATTACHMENT | HBC | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | MOTOR DEVICE |