XMAX MOTOR
Report
- Report Number
- 1045834-2014-12811
- Event Type
- Malfunction
- Date Received
- September 15, 2014
- Date of Event
- August 29, 2014
- Report Date
- September 2, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- ERL
- PMA / PMN Number
- PK965080
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
THE ACTUAL DEVICE WAS RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE RELIABILITY ENGINEERING EVALUATES THE DEVICE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS NOT CONFIRMED. AN ASSESSMENT WAS PERFORMED AND THE DEVICE MET ALL MANUFACTURE'S SPECIFICATIONS. THEREFORE, AN ASSIGNABLE ROOT CAUSE WAS NOT DETERMINED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
IT WAS REPORTED THAT DURING STERILE PROCESSING, IT WAS OBSERVED THAT THE HOSE ON THE MOTOR DEVICE HAD A PIN HOLE AND AN AIR LEAK. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 570132 | XMAX MOTOR | DRILL, SURGICAL, ENT (PNEUMATIC) - HANDPIECE | ERL | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |