FDA Adverse Event Malfunction Summary report: N

IMRI SHORT HD ATTACHMENT

MDR report key: 4090618 · Received September 15, 2014

Report

Report Number
1045834-2014-12867
Event Type
Malfunction
Date Received
September 15, 2014
Date of Event
August 28, 2014
Report Date
August 28, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBB
PMA / PMN Number
PK061297
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE PASSED ALL OPERATIONAL SPECIFICATIONS. THEREFORE, THE REPORTED CONDITION COULD NOT BE CONFIRMED. AN ASSIGNABLE ROOT CAUSE WAS NOT DETERMINED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE RELIABILITY ENGINEERING EVALUATES THE DEVICE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

THIS IS REPORT 2 OF 2 OF THE SAME EVENT IT WAS REPORTED THAT DURING CRANIOTOMY SURGERY, IT WAS OBSERVED THAT THE MOTOR DEVICE AND THE ATTACHMENT DEVICE WOULD NOT LOAD THE BURR DEVICE CORRECTLY WHEN IN USE TOGETHER. IT WAS FURTHER CLARIFIED THAT THE BURR DEVICE WOULD NOT LOCK INTO PLACE. AS A RESULT, THERE WAS A TEN MINUTE EXTENSION TO THE SURGICAL PROCEDURE. A SPARE DEVICE WAS AVAILABLE FOR USE AND THE SURGICAL PROCEDURE WAS WHICH COMPLETED SUCCESSFULLY WITH A SPARE DEVICE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
570202 IMRI SHORT HD ATTACHMENT MOTOR, DRILL, PNEUMATIC HBB DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1 MOTOR DEVICE| BURR DEVICE