FDA Adverse Event Malfunction Summary report: N

ELECTRIC PEN DRIVE

MDR report key: 4090615 · Received September 15, 2014

Report

Report Number
2520274-2014-13666
Event Type
Malfunction
Date Received
September 15, 2014
Date of Event
August 21, 2014
Report Date
August 26, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HWE
PMA / PMN Number
PK043310
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS NOT CONFIRMED. A FUNCTIONAL ASSESSMENT WAS PERFORMED AND THE DEVICE MEETS ALL MANUFACTURER'S SPECIFICATIONS. THEREFORE, AN ASSIGNABLE ROOT CAUSE WAS NOT DETERMINED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Additional Manufacturer Narrative · 1

THE MANUFACTURING LOCATION WAS UNKNOWN. THE DEVICE MANUFACTURE DATE IS UNKNOWN. AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE ELECTRIC PEN DRIVE DEVICE WOULD NOT RUN IN FORWARD BUT WOULD WORK IN REVERSE. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE. A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
569681 ELECTRIC PEN DRIVE INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE DEPUY SYNTHES POWER TOOLS AV21787/2

Patients

Seq Age Sex Outcome Treatment