FDA Adverse Event Malfunction Summary report: N

1/8" PEG DRILL

MDR report key: 4090568 · Received September 15, 2014

Report

Report Number
0002249697-2014-03470
Event Type
Malfunction
Date Received
September 15, 2014
Date of Event
August 21, 2014
Report Date
August 21, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
MBH
PMA / PMN Number
K123486
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

WHEN COMPLETED, THE INVESTIGATION RESULTS WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

AN EVENT REGARDING A FRACTURED TRIATHLON 1/8" PEG DRILL WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD & RESULTS: -DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION CONFIRMED THE REPORTED FRACTURE. A MATERIAL ANALYSIS INDICATED THE DRILL BIT FRACTURE OCCURRED IN MIXED-MODE OVERLOAD AND DUCTILE OVERLOAD WHILE IN RIGHT-HANDED ROTATION. -MEDICAL RECORDS RECEIVED AND EVALUATION: NOT PERFORMED AS THERE IS NO INDICATION PATIENT FACTORS CONTRIBUTED TO THE REPORTED EVENT. -DEVICE HISTORY REVIEW: THE REPORTED DEVICE WAS MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. -COMPLAINT HISTORY REVIEW: THERE HAS BEEN ONE OTHER SIMILAR EVENTS REPORTED FOR THIS MANUFACTURING LOT. CONCLUSIONS: THE EVENT WAS CONFIRMED, THE DRILL BIT FRACTURED. A MATERIAL ANALYSIS INDICATED THE DRILL BIT FRACTURE OCCURRED IN MIXED-MODE OVERLOAD AND DUCTILE OVERLOAD WHILE IN RIGHT-HANDED ROTATION.

Description of Event or Problem · 1

DRILL BIT BROKE DURING USE.

Description of Event or Problem · 1

DRILL BIT BROKE DURING USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
569601 1/8" PEG DRILL INSTRUMENT MBH STRYKER ORTHOPAEDICS-MAHWAH SC4C07

Patients

Seq Age Sex Outcome Treatment
1 Other