FDA Adverse Event Malfunction Summary report: N

STERRAD NX STERILIZER

MDR report key: 4090521 · Received September 15, 2014

Report

Report Number
2084725-2014-00308
Event Type
Malfunction
Date Received
September 15, 2014
Date of Event
September 5, 2014
Report Date
September 5, 2014
Manufacturer
ADVANCED STERILIZATION PRODUCTS
Product Code
MLR
PMA / PMN Number
K042116
Removal / Correction Number
Z-0744-0746-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER PERFORMED A PREVENTATIVE MAINTENANCE (PM1) AND THE VACUUM PUMP OIL, CATALYTIC DECOMP FILTER AND OIL MIST FILTER WERE REPLACED. UNIT MEETS SPECIFICATIONS AND WAS RETURNED TO SERVICE ON (B)(4) 2014.

Additional Manufacturer Narrative · 1

MANUFACTURER DATE: 10/30/2007. CONCLUSION: BASED ON CAPA INVESTIGATION AND ASTM TESTING, OIL DEGRADATION CAN CAUSE ODOR/SMELLS FOR STERRAD SYSTEMS. ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD (DHR), SERVICE HISTORY, TRENDING OF THE PRODUCT MALFUNCTION CODE, FAILURE MODE AND EFFECTS ANALYSIS (FMEA), AND SYSTEM HAZARD AND USER MISUSE ANALYSIS (SHUMA), AND CORRECTIVE AND PREVENTATIVE ACTION (CAPA). THE DHR WAS REVIEWED. THE UNIT MET MANUFACTURING SPECIFICATIONS AT THE TIME OF RELEASE AND THERE WERE NO ISSUES RELATED TO THIS FAILURE MODE NOTED. THE SERVICE HISTORY FOR THIS UNIT FOR THE PAST SIX MONTHS (03/13/2014 THROUGH 09/09/2014) DID NOT OBSERVE ANY SIGNIFICANT TREND. THE TREND FOR THE PRODUCT MALFUNCTION CODE OF ODOR/SMELLS WAS COMPLETED FROM OCTOBER 2013 THROUGH SEPTEMBER 2014 AND REVEALED A SIGNIFICANT TREND WHICH IS ADDRESSED IN CAPA. THE FMEA REVEALED THE RISK PRIORITY NUMBER (RPN) SCORES ARE CONSIDERED TO BE ACCEPTABLE. THE SHUMA INDICATES THE RISK IS AS LOW AS REASONABLY PRACTICABLE FOR MILD (LIMITED) EXPOSURE TO ODOR OR ODORANTS. CAPA INVESTIGATION AS WELL AS THE ASTM TESTING PERFORMED INDICATED OIL DEGRADATION CAN CAUSE ODOR/SMELLS FOR STERRAD® SYSTEMS. NO PARTS WERE RETURNED FOR FURTHER EVALUATION. THE REPORTED ISSUE WAS RESOLVED AT THE CUSTOMER FACILITY. THE ISSUE WILL BE TRACKED AND TRENDED. NO FURTHER INVESTIGATION IS NECESSARY AT THIS TIME.

Description of Event or Problem · 1

WHILE ONSITE SERVICING THE STERRAD® NX STERILIZER FOR ANOTHER ISSUE, A FIELD SERVICE ENGINEER (FSE) DISCOVERED AN ODOR EMITTING FROM THE UNIT. THERE WAS NO REPORT OF HUMAN REACTION. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(6) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
569639 STERRAD NX STERILIZER STERRAD EQUIPMENT (MLR) MLR ADVANCED STERILIZATION PRODUCTS NA 0033070796

Patients

Seq Age Sex Outcome Treatment
1