HOMECHOICE

MDR report key: 4090492 · Received September 15, 2014

Report

Report Number
1416980-2014-31332
Event Type
Malfunction
Date Received
September 15, 2014
Date of Event
April 27, 2014
Report Date
August 21, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS EVALUATED AND THE ISSUE WAS IDENTIFIED DURING THE REVIEW OF THE EVENT HISTORY LOGS. THE CAUSE WAS DETERMINED TO BE USE ERROR, INAPPROPRIATE BYPASS OF THE DRAIN, NOT INCLUDING INITIAL DRAIN. HOMECHOICE / HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE GIVES INSTRUCTIONS ON HOW TO BYPASS THE DRAIN PHASE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 04:00:19. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 1509ML, INDICATING THE HOME PATIENT (HP) DRAINED 1509ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2200ML. NO ADDITIONAL INFORMATION IS AVAILABLE.