FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4090438 · Received September 15, 2014

Report

Report Number
2032227-2014-23126
Event Type
Malfunction
Date Received
September 15, 2014
Date of Event
August 16, 2014
Report Date
August 16, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH MOISTURE DAMAGE AT ELECTRONIC ASSEMBLY. THE DEVICE WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP, AND BROKEN BELT CLIP SLOT.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THEY JUMPED INTO THE POOL WITH THE INSULIN PUMP ON. CUSTOMER STATED THAT FLUID WAS VISIBLE UNDER THE DISPLAY OF THE DEVISE. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 176 MG/DL. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
568638 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 24 YR