FDA Adverse Event Malfunction Summary report: N

OHMEDA

MDR report key: 409039 · Received August 2, 2002

Report

Report Number
MW1025754
Event Type
Malfunction
Date Received
August 2, 2002
Report Date
August 2, 2002
Manufacturer
*
Product Code
LBI
Product Problem
Yes
Report Source
Voluntary report
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

ADD'L INFO REC'D FROM MFR 9/16/02: THE PRODUCT PHOTOTHERAPY LIGHTS MENTIONED ARE NOT SOLD BY RH BURTON CO. NOT SURE WHAT THIS PRODUCT IS, BUT THERE IS ANOTHER BURTON MEDICAL PRODUCTS CO IN ANOTHER STATE.

Description of Event or Problem · 1

WHILE CONDUCTING A RANDOMIZED CLINICAL TRIAL OF AGGRESSIVE VS CONSERVATIVE PHOTOTHERAPY IN PREMATURE INFANTS, A PHYSICIAN WAS ALARMED BY HIGH UVA READINGS THAT WERE OBTAINED FROM THE PHOTOTHERAPY SPOTLIGHTS. THE SPOTLIGHT READINGS ARE ORDERS OF MAGNITUDE HIGHER THAN THE CURRENT IEC NEONATAL PHOTOTHERAPY STANDARD (1X10 TO NEGATIVE 2 MICROWATTS) AND COULD BE HARMFUL TO INFANTS. ABOUT HALF OF THE NEONATAL UNITS IN THE STUDY USE THESE SPOTLIGHTS AS THE MAIN SOURCE OF PHOTOTHERAPY. THESE LIGHTS ARE WIDELY USED ACROSS THE COUNTRY AND INTERNATIONALLY. THE PHYSICIAN REQUESTS THAT THE FDA DETERMINE IF THIS IS A DANGER. SUMMARY OF THE PHYSICIAN'S FINDINGS: SPOTLIGHTS: OHMEDA, DISTANCE: 78CM, UVA METER: 22.5. DISTANCE: 43CM UVA METER 69, IRRADIANCE: 16.7, UVA METER: 34. DISTANCE 72 CM, IRRADIANCE: 9.7, UVA METER: 42, DISTANCE: 55 CM, IRRADIANCE: 16.5, UVA METER: 79, DISTANCE: 45 CM, IRRADIANCE: 30, UVA METER: 125, DISTANCE: 35 CM, IRRADIANCE: 39, UVA METER: 224. SPOTLIGHTS: BURTON HALOGEN, IRRADIANCE: 20, UVA METER: 70, DISTANCE: ISOLETTE, IRRADIANCE: 20, UVA METER: 0.02.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OHMEDA SPOTLIGHT LBI * NI NI
2 BURTON HALOGEN SPOTLIGHT LBI * NI NI

Patients

Seq Age Sex Outcome Treatment
1 NO INFO