5MM, 33CM PEEK MONOPOLAR HANDLE 33CM
Report
- Report Number
- 0002936485-2014-00713
- Event Type
- Malfunction
- Date Received
- September 15, 2014
- Date of Event
- August 25, 2014
- Report Date
- August 25, 2014
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- KOG
- PMA / PMN Number
- K973259
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. DURING THE INSPECTION OF THE 5MM, 33CM PEEK MONOPOLAR HANDLE 33CM IT WAS CONFIRMED, THE INSULATION IS COMPROMISED. VISIBLE DINGS AND SCRATCHES THROUGHOUT THE SHAFT AND SEPARATION OF THE INSULATION AT THE PROXIMAL AND DISTAL END WERE SEEN. ALSO IT WAS NOTICED THAT THIS UNIT COULD HAVE POSSIBLY BEEN REPAIRED BY A THIRD PARTY IN THE PAST. THE 5MM, 33CM PEEK MONOPOLAR HANDLE 33CMIS MISSING THE STRYKER ETCHING AT THE PROXIMAL END OF THE SHAFT, WHICH CONSIST OF LOT NUMBER, CE MARK, US PAT NO AND THE REFERENCE NUMBER. THE 5MM, 33CM PEEK MONOPOLAR HANDLE 33CM FAILED THE INSULATION INTEGRITY TEST. IFU 1000401070 STATES "BEFORE USE, ENSURE THAT THERE ARE NO BREAKS, CHIPS, CRACKS, SCRATCHES, TEARS, OR MISSING/LOOSE COMPONENTS ON THE SHAFT INSULATION, HANDLE, OR HOUSING. DO NOT USE THE INSTRUMENT IF ANY OF THESE DEFECTS ARE PRESENT AS THIS COULD CAUSE UNINTENDED ELECTROSURGICAL BURNS AND LIFE THREATENING COMPLICATIONS. IF DAMAGE HAS OCCURRED, DISCONTINUE USE AND RETURN THE INSTRUMENT FOR REPAIR OR REPLACEMENT." THE PROBABLE ROOT CAUSE COULD BE WEAR AND TEAR AND OR MISHANDLING. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 568964 | 5MM, 33CM PEEK MONOPOLAR HANDLE 33CM | ENDOSCOPE AND/OR ACCESSORIES | KOG | STRYKER ENDOSCOPY-SAN JOSE | 08252014 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |