FDA Adverse Event Malfunction Summary report: N

5MM, 33CM PEEK MONOPOLAR HANDLE 33CM

MDR report key: 4090353 · Received September 15, 2014

Report

Report Number
0002936485-2014-00713
Event Type
Malfunction
Date Received
September 15, 2014
Date of Event
August 25, 2014
Report Date
August 25, 2014
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
KOG
PMA / PMN Number
K973259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. DURING THE INSPECTION OF THE 5MM, 33CM PEEK MONOPOLAR HANDLE 33CM IT WAS CONFIRMED, THE INSULATION IS COMPROMISED. VISIBLE DINGS AND SCRATCHES THROUGHOUT THE SHAFT AND SEPARATION OF THE INSULATION AT THE PROXIMAL AND DISTAL END WERE SEEN. ALSO IT WAS NOTICED THAT THIS UNIT COULD HAVE POSSIBLY BEEN REPAIRED BY A THIRD PARTY IN THE PAST. THE 5MM, 33CM PEEK MONOPOLAR HANDLE 33CMIS MISSING THE STRYKER ETCHING AT THE PROXIMAL END OF THE SHAFT, WHICH CONSIST OF LOT NUMBER, CE MARK, US PAT NO AND THE REFERENCE NUMBER. THE 5MM, 33CM PEEK MONOPOLAR HANDLE 33CM FAILED THE INSULATION INTEGRITY TEST. IFU 1000401070 STATES "BEFORE USE, ENSURE THAT THERE ARE NO BREAKS, CHIPS, CRACKS, SCRATCHES, TEARS, OR MISSING/LOOSE COMPONENTS ON THE SHAFT INSULATION, HANDLE, OR HOUSING. DO NOT USE THE INSTRUMENT IF ANY OF THESE DEFECTS ARE PRESENT AS THIS COULD CAUSE UNINTENDED ELECTROSURGICAL BURNS AND LIFE THREATENING COMPLICATIONS. IF DAMAGE HAS OCCURRED, DISCONTINUE USE AND RETURN THE INSTRUMENT FOR REPAIR OR REPLACEMENT." THE PROBABLE ROOT CAUSE COULD BE WEAR AND TEAR AND OR MISHANDLING. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.