FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 4090341 · Received September 15, 2014

Report

Report Number
2032227-2014-19181
Event Type
Malfunction
Date Received
September 15, 2014
Date of Event
March 4, 2013
Report Date
August 19, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
LZG
PMA / PMN Number
K040676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH UNRESPONSIVE BUTTONS AS A RESULT OF CORRODED KEYPAD TRACES. UNABLE TO CONFIRM THE ALARM. THE DEVICE WAS RECEIVED WITH SCRATCHED SCREEN, CRACKED RESERVOIR TUBE LIP AND TUBE THREADS AS WELL WITH MISSING END CAP STICKER. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP HAD A BUTTON ERROR ALARM AND THE ESCAPE AND ACTIVATE BUTTONS WERE NOT RESPONDING. ADVISED THE CALLER TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
568960 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-715NAL

Patients

Seq Age Sex Outcome Treatment
1