FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 4090115 · Received July 17, 2014

Report

Report Number
1824206-2014-01990
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
July 11, 2014
Report Date
July 11, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K962942
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE SIDE RAIL LATCH ASSEMBLY WAS STICKY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2013. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN CLEANED AND LUBRICATED THE SIDE RAIL LATCH ASSEMBLY TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE SIDE RAIL WOULD NOT LATCH. THE BED WAS LOCATED IN THE ICU. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).