FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION WITH TELEMETRY

MDR report key: 4090091 · Received August 18, 2014

Report

Report Number
2221819-2014-00500
Event Type
Malfunction
Date Received
August 18, 2014
Date of Event
July 14, 2014
Report Date
July 16, 2014
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
031760
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPANY REPRESENTATIVE EVALUATED THE SYSTEM. CORRECTIONS INCLUDED REPLACEMENT OF THE WIRELESS TRANSCEIVER'S POWER SUPPLY AND ADJUSTING THE PANEL ANTENNA. SYSTEM WAS CALIBRATED AND SAFETY TESTED TO FACTORY'S SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED AN ISSUE WITH THE PANORAMA CENTRAL STATION, WHICH MAY HAVE AFFECTED TELEMETRY MONITORING. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
496248 PANORAMA CENTRAL STATION WITH TELEMETRY CENTRAL MONITORING STATION MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1