FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION WITH TELEMETRY

MDR report key: 4090087 · Received August 18, 2014

Report

Report Number
2221819-2014-00495
Event Type
Malfunction
Date Received
August 18, 2014
Date of Event
July 2, 2014
Report Date
July 11, 2014
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
031760
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPANY REPRESENTATIVE EVALUATED THE SYSTEM. CORRECTIONS INCLUDED CLEANING THE UNITS AND REPLACEMENTS OF THE RIBBON CABLE, INTERFACE PCB. UNITS AND SYSTEM WERE CALIBRATED AND SAFETY TESTED TO FACTORY'S SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED ISSUES RELATED THE PANORAMA CENTRAL STATION'S AMBULATORY TELEPACKS, WHICH MAY HAVE AFFECTED TELEMETRY MONITORING. NO PATIENT INJURY WAS REPORTED. AFFECTED SERIAL #S: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
496270 PANORAMA CENTRAL STATION WITH TELEMETRY CENTRAL MONITORING STATION MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1