FDA Adverse Event
Malfunction
Summary report: N
ACCUTORR V MONITOR
MDR report key: 4090073
·
Received August 18, 2014
Report
- Report Number
- 2221819-2014-00481
- Event Type
- Malfunction
- Date Received
- August 18, 2014
- Date of Event
- June 19, 2014
- Report Date
- July 8, 2014
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- MHX
- PMA / PMN Number
- 091068
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
COMPANY REP EVALUATED THE UNIT. CORRECTIONS INCLUDED REPLACEMENTS OF THE UNIT'S SPO2 BOARD AND LCD PANEL. UNIT WAS CALIBRATED AND WAS TESTED TO FACTORY'S SPECS.
Description of Event or Problem · 1
CUSTOMER REPORTED AN ISSUE WITH ACCUTORR V MONITOR, WHICH MAY HAVE AFFECTED SPO2 MONITORING. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 496306 | ACCUTORR V MONITOR | PATIENT MONITOR | MHX | MINDRAY DS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |