FDA Adverse Event Malfunction Summary report: N

IC PROFILER-MR

MDR report key: 4084960 · Received September 3, 2014

Report

Report Number
1038814-2014-00001
Event Type
Malfunction
Date Received
September 3, 2014
Date of Event
August 6, 2014
Report Date
August 6, 2014
Manufacturer
SUN NUCLEAR CORP.
Product Code
LHO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE IC PROFILER-MR IS NOT USED IN THE PRESENCE OF A PATIENT. HOWEVER, STAFF SETTING UP THE IC PROFILER-MR FOR MEASUREMENT OF AN INSTALLED VIEWRAY DELIVERY SYSTEM COULD EXPERIENCE A PINCH HAZARD BETWEEN THE COVERS ON THE DELIVERY SYSTEM AND THE IC PROFILER-MR IF IT IS ATTRACTED TOWARDS THE DELIVERY SYSTEM BY THE MAGNETIC FIELD. INVESTIGATION IDENTIFIED AN ASSEMBLY ERROR OF THIS UNIT. THERE WERE 3 UNITS MANUFACTURED IN THE LOT AND 9 UNITS MANUFACTURED IN TOTAL. AS PART OF THE INVESTIGATION OF THIS INCIDENT, ALL UNITS MANUFACTURED HAVE BEEN EVALUATED. DATA OBTAINED CONFIRMED 2 UNITS CONTAINED THE ASSEMBLY ERROR. THESE UNITS HAVE BEEN REMOVED AND REPAIRED.

Description of Event or Problem · 1

STAFF USING AN IC PROFILER-MR AT (B)(6) RADIATION ONCOLOGY REPORTED THEIR IC PROFILER-MR UNIT THAT WAS RESTING ON THE COUCH MOVED WHEN THE COUCH WAS TRANSLATED INTO A MAGNETIC FIELD OF A VIEWRAY DELIVERY SYSTEM. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
535526 IC PROFILER-MR IC PROFILER-MR LHO SUN NUCLEAR CORP. 1123 W92327

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention VIEWRAY SYSTEM