IC PROFILER-MR
Report
- Report Number
- 1038814-2014-00001
- Event Type
- Malfunction
- Date Received
- September 3, 2014
- Date of Event
- August 6, 2014
- Report Date
- August 6, 2014
- Manufacturer
- SUN NUCLEAR CORP.
- Product Code
- LHO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE IC PROFILER-MR IS NOT USED IN THE PRESENCE OF A PATIENT. HOWEVER, STAFF SETTING UP THE IC PROFILER-MR FOR MEASUREMENT OF AN INSTALLED VIEWRAY DELIVERY SYSTEM COULD EXPERIENCE A PINCH HAZARD BETWEEN THE COVERS ON THE DELIVERY SYSTEM AND THE IC PROFILER-MR IF IT IS ATTRACTED TOWARDS THE DELIVERY SYSTEM BY THE MAGNETIC FIELD. INVESTIGATION IDENTIFIED AN ASSEMBLY ERROR OF THIS UNIT. THERE WERE 3 UNITS MANUFACTURED IN THE LOT AND 9 UNITS MANUFACTURED IN TOTAL. AS PART OF THE INVESTIGATION OF THIS INCIDENT, ALL UNITS MANUFACTURED HAVE BEEN EVALUATED. DATA OBTAINED CONFIRMED 2 UNITS CONTAINED THE ASSEMBLY ERROR. THESE UNITS HAVE BEEN REMOVED AND REPAIRED.
STAFF USING AN IC PROFILER-MR AT (B)(6) RADIATION ONCOLOGY REPORTED THEIR IC PROFILER-MR UNIT THAT WAS RESTING ON THE COUCH MOVED WHEN THE COUCH WAS TRANSLATED INTO A MAGNETIC FIELD OF A VIEWRAY DELIVERY SYSTEM. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 535526 | IC PROFILER-MR | IC PROFILER-MR | LHO | SUN NUCLEAR CORP. | 1123 | W92327 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | VIEWRAY SYSTEM |