OLYMPUS DUODENOVIDEOSCOPE
Report
- Report Number
- 2951238-2014-00359
- Event Type
- Injury
- Date Received
- August 22, 2014
- Date of Event
- January 1, 2013
- Report Date
- July 25, 2014
- Manufacturer
- OLYMPUS MEDICAL SYSTEM CORP
- Product Code
- KOG
- PMA / PMN Number
- K024033
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
NO DEVICES WERE RETURNED TO OLYMPUS FOR EVAL. OLYMPUS OFFERED AN ON-SITE VISIT TO THE USER FACILITY TO ASSESS THEIR REPROCESSING PRACTICES, BUT THE USER FACILITY DECLINED. THE EXACT CAUSE OF THE PT OUTCOME COULD NOT BE CONCLUSIVELY DETERMINED. IF ADD'L AND SIGNIFICANT INFO BECOMES AVAILABLE AT A LATER TIME, THIS REPORT WILL BE SUPPLEMENTED. CROSS REFERENCE MFR REPORT NUMBERS: 2951238-2014-00348, 2951238-2014-00349, 2951238-2014-00350, 2951238-2014-00350, 2951238-2014-00351, 2951238-2014-00352, 2951238-2014-00353, 2951238-2014-00354, 2951238-2014-00355, 2951238-2014-00356, 2951238-2014-00357, 2951238-2014-00358, 2951238-2014-00360, 2951238-2014-00361, 2951238-2014-00362, 2951238-2014-00363, 2951238-2014-00364, 2951238-2014-00365, 2951238-2014-00366, 2951238-2014-00367, 2951238-2014-00368, 2951238-2014-00369, 2951238-2014-00370, 2951238-2014-00371, 2951238-2014-00372, 2951238-2014-00373, 2951238-2014-00374, 2951238-2014-00375, 2951238-2014-00376, 2951238-2014-00377, 2951238-2014-00378, 2951238-2014-00379, 2951238-2014-00380, 2951238-2014-00381, 2951238-2014-00382, 2951238-2014-00383.
OLYMPUS FOLLOWED UP WITH THE USER FACILITY MULTIPLE TIMES VIA TELEPHONE AND IN WRITING TO OBTAIN ADDITIONAL INFO REGARDING THE REPORTED EVENTS, BUT WAS UNSUCCESSFUL. THE INITIAL FOLLOW-UP WAS MADE ON JULY 28, 2014, VIA TELEPHONE AND IN WRITING. SUBSEQUENT FOLLOW-UPS WERE MADE ON THE FOLLOWING DATES: JULY 29, 2014 VIA ELECTRONIC MAIL, JULY 30, 2014, AUGUST 7, 2014 VIA A TELECONFERENCE CALL WITH ATTENDEES FROM THE USER FACILITY AND OLYMPUS, AUGUST 12, 2014, AUGUST 14, 2014, AUGUST 20, 21, 22, 2014. ON AUGUST 27, 2014, THE USER FACILITY DECLINED THE ON-SITE VISIT THAT OLYMPUS OFFERED IN AN ATTEMPT TO COLLABORATE ON A PHYSICAL INSPECTION OF ALL THE DUODENOSCOPES INVOLVED IN THE REPORTED EVENTS. OLYMPUS PERFORMED AN INVENTORY SEARCH FOR DEVICES SOLD TO THE USER FACILITY BASED ON THE NEWS ARTICLE DATED JANUARY 21, 2015 AND DETERMINED THE USER FACILITY HAD FIVE (5) TJF-160VF AND 23 TJF-Q180V. BASED ON THE HIGHER PROBABILITY THAT A 180 MODEL WAS USED ON THE REPORTED EVENTS, OLYMPUS SUPPLEMENTED THE 37 MDRS CHANGING THE DUODENOSCOPE MODEL FROM TJF-160VF TO TJF-Q180V. HOWEVER, THE SPECIFIC MODEL OF DUODENOSCOPES USED DURING THE REPORTED EVENTS STILL REMAINS UNK, AS THE USER FACILITY HAS NOT RESPONDED TO OLYMPUS' INQUIRIES.
OLYMPUS WAS INFORMED OF 37 ALLEGED INFECTIONS AT THE USER FACILITY INVOLVING MULTIPLE DUODENOSCOPES (160 SERIES/180SERIES). IN EARLY 2013 THE WASHINGTON STATE DEPT OF HEALTH IDENTIFIED SEVEN CASES OF CARBAPENEM RESISTANT ENTEROBACTERIACEAE (CRE) CULTURES FROM SAMPLES SENT BY THE USER FACILITY. THERE WERE NO SPECIFIC ENDOSCOPE MODELS OR SERIAL NUMBERS PROVIDED BY THE USER FACILITY. FOUR DEATHS WERE IDENTIFIED BY THE USER FACILITY. THE CAUSE OF PT DEATHS IS UNK. ADDITIONALLY, OLYMPUS WAS NOTIFIED OF 30 PT INFECTIONS BASED ON CULTURE OF VARIANT E.COLI FROM PT'S BLOOD, BILE, URINE, OR THE RESPIRATORY TRACT. OLYMPUS CONTACTED THE USER FACILITY VIA TELEPHONE REGARDING THE REPORTED EVENTS BUT NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 508317 | OLYMPUS DUODENOVIDEOSCOPE | DUODENOVIDEOSCOPE | KOG | OLYMPUS MEDICAL SYSTEM CORP | TJF-160VF | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | TJF-Q180V SERIAL NUMBER (UNK)| TJF-Q180V SERIAL NUMBER (UNK) |