FDA Adverse Event
Malfunction
Summary report: N
OT PING ENHANCED METER
MDR report key: 4083400
·
Received September 12, 2014
Report
- Report Number
- 2939301-2014-23991
- Event Type
- Malfunction
- Date Received
- September 12, 2014
- Report Date
- September 8, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FOLLOW-UP # 1 ¿ (11/12/2014). THE PATIENT¿S METER HAS BEEN RETURNED ON 10/1/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 10/30/2014 WITH THE FOLLOWING FINDINGS:THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER¿S LCD WAS FOUND TO BE DEFECTIVE. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
Description of Event or Problem · 1
ON (B)(6)2014, THE REPORTER CONTACTED LIFESCAN (B)(6), ALLEGING FADING DISPLAY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 564884 | OT PING ENHANCED METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |