PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-21355
- Event Type
- Malfunction
- Date Received
- September 11, 2014
- Date of Event
- August 13, 2014
- Report Date
- August 13, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP ALARMED MOTOR ERROR DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNABLE TO PERFORM FUNCTIONAL TESTING DUE TO MOTOR ERROR ALARMS. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHED LCD WINDOW.
IT IS REPORTED THAT A CUSTOMER RECEIVED A MOTOR ERROR ALARM ON THEIR INSULIN PUMP. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS AT 87 MG/DL. THE CUSTOMER STATED THAT THEY WERE EXPOSED TO AN MRI WHICH COULD HAVE LED TO THE ISSUE. THE CUSTOMER WAS INFORMED THAT THE PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK UP PLAN. A NEW PUMP WAS SHIPPED TO THE CUSTOMER. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 560625 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723RNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |