FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4082641 · Received September 11, 2014

Report

Report Number
2032227-2014-21476
Event Type
Malfunction
Date Received
September 11, 2014
Date of Event
August 13, 2014
Report Date
August 13, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD UNRESPONSIVE UP AND DOWN ARROW BUTTONS DUE TO CORRODED KEYPAD TRACES. NO ALARM FOR AN UNEXPECTED RESTART COULD BE VERIFIED DUE TO THE UNRESPONSIVE BUTTONS. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP WAS ALARMING. CUSTOMER STATED THE DEVICE SHUTS DOWN AND THEN TURNS ON WHILE NOT ALLOWING ANY USER INPUTS. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 160 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
561951 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 36 YR