FDA Adverse Event
Malfunction
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR
MDR report key: 4081443
·
Received January 14, 2014
Report
- Report Number
- 2938836-2014-06174
- Event Type
- Malfunction
- Date Received
- January 14, 2014
- Date of Event
- April 16, 2012
- Manufacturer
- ST. JUDE MEDICAL INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFO WAS PROVIDED BY MFR, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PT RECEIVED A VIBRATORY ALERT. UPON INTERROGATION, HIGH IMPEDANCE WAS OBSERVED. DURING TESTING, THE HIGH MEASUREMENT COULD NOT BE REPRODUCED. IMPEDANCE MEASUREMENTS WERE NORMAL. PT WOULD BE MONITORED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 35065 | DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL INC., CRMD | 7120Q/58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |