FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4081432 · Received September 11, 2014

Report

Report Number
2032227-2014-21268
Event Type
Malfunction
Date Received
September 11, 2014
Date of Event
July 5, 2014
Report Date
August 12, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY TEST. NO MOTOR ERROR ALARM NOTED. MOTOR TEST AND REWIND FUNCTIONED PROPERLY DURING TESTING. UNIT RECEIVED WITH CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED DISPLAY WINDOW, AND RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

IT IS REPORTED THAT A CUSTOMER RECEIVED A MOTOR ERROR ALARM ON THEIR INSULIN PUMP. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS AT 187 MG/DL. THE CUSTOMER STATED THAT THERE WERE NO SIGNIFICANT EVENTS THAT COULD HAVE LED TO THE ISSUE. THE CUSTOMER WAS INFORMED THAT THE PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK UP PLAN. A NEW PUMP WAS SHIPPED TO THE CUSTOMER. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
562552 530G INSULIN PUMP OZO OZO MEDTRONIC MINIMED MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 61 YR