FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 4080967 · Received July 9, 2014

Report

Report Number
3008642652-2014-02129
Event Type
Malfunction
Date Received
July 9, 2014
Date of Event
May 21, 2014
Report Date
July 7, 2014
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHARGER/MODEM WON'T POWER UP) HAS BEEN CONFIRMED. UPON INVESTIGATION THE BATTERY CHARGER WAS UNABLE TO POWER ON. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A DEFECTIVE POWER SUPPLY BRICK. THE ROOT CAUSE FOR THE DEFECTIVE POWER BRICK COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE POWER SUPPLY BRICK. THE PT RECEIVED A REPLACEMENT BATTERY CHARGER/MODEM.

Description of Event or Problem · 1

A PT SERVICE REPRESENTATIVE (PSR) CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A CHARGER/MODEM THAT WOULD NOT POWER UP. THE PSR WAS SENT A REPLACEMENT CHARGER/MODEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
400954 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA