FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 4080966
·
Received July 9, 2014
Report
- Report Number
- 3008642652-2014-02104
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Date of Event
- May 23, 2014
- Report Date
- July 3, 2014
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (EXPOSED WIRES) HAS BEEN CONFIRMED. UPON INVESTIGATION, THE CABLE CONNECTING THE DISTRIBUTION NODE TO THE REAR THERAPY ELECTRODES WAS PULLED FROM THE STRAIN RELIEF, DAMAGING WIRES. THE ROOT CAUSE FOR THE STRAINED CABLE COULD NOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE STRAINED CABLE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) YEAR OLD MALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT WIRES WERE EXPOSED ON HIS ELECTRODE BELT. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 400872 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR |