NOVOSYN VIOLET 2/0 (3) 6X45CM
Report
- Report Number
- 2916714-2014-00686
- Event Type
- Malfunction
- Date Received
- August 27, 2014
- Date of Event
- July 21, 2014
- Report Date
- August 27, 2014
- Manufacturer
- B. BRAUN SURGICAL S.A..
- Product Code
- GAM
- PMA / PMN Number
- K122734
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
US REPORTING AGENT NOTIFIED ON: (B)(4) 2014. MFG SITE EVAL: SAMPLES RECEIVED: 32 UNOPENED POUCHES. THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE/BATCH. THERE ARE UNITS IN OEM STOCK. TESTED THE KNOT SECURITY CONTROL OF THE SAMPLES RECEIVED AND THE RESULTS ARE INTO THE CURRENT RANGE FOR THIS THREAD AND SIZE. TESTED THE KNOT PULL TENSILE STRENGTH OF THE SAMPLES RECEIVED AND THE RESULTS FULFILL THE REQUIREMENTS OF THE OEM. REVIEWED THE BATCH MFG RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFILL OEM REQUIREMENTS. FINAL CONCLUSION: THE COMPLAINT IS NOT CORRESPONDING (NOT JUSTIFIED). ACTIONS ON PRODUCT: NOT APPLICABLE. CORRECTIVE/PREVENTIVE ACTIONS: NOT APPLICABLE.
COUNTY OF COMPLAINT: (B)(6). COMPLAINT STATES: KNOT SECURITY IS DECAYED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 518794 | NOVOSYN VIOLET 2/0 (3) 6X45CM | SYNTHETIC ABSORBABLE SUTURE | GAM | B. BRAUN SURGICAL S.A.. | C0058646 | 114223 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |