FDA Adverse Event Malfunction Summary report: N

NOVOSYN VIOLET 2/0 (3) 6X45CM

MDR report key: 4080787 · Received August 27, 2014

Report

Report Number
2916714-2014-00686
Event Type
Malfunction
Date Received
August 27, 2014
Date of Event
July 21, 2014
Report Date
August 27, 2014
Manufacturer
B. BRAUN SURGICAL S.A..
Product Code
GAM
PMA / PMN Number
K122734
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

US REPORTING AGENT NOTIFIED ON: (B)(4) 2014. MFG SITE EVAL: SAMPLES RECEIVED: 32 UNOPENED POUCHES. THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE/BATCH. THERE ARE UNITS IN OEM STOCK. TESTED THE KNOT SECURITY CONTROL OF THE SAMPLES RECEIVED AND THE RESULTS ARE INTO THE CURRENT RANGE FOR THIS THREAD AND SIZE. TESTED THE KNOT PULL TENSILE STRENGTH OF THE SAMPLES RECEIVED AND THE RESULTS FULFILL THE REQUIREMENTS OF THE OEM. REVIEWED THE BATCH MFG RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFILL OEM REQUIREMENTS. FINAL CONCLUSION: THE COMPLAINT IS NOT CORRESPONDING (NOT JUSTIFIED). ACTIONS ON PRODUCT: NOT APPLICABLE. CORRECTIVE/PREVENTIVE ACTIONS: NOT APPLICABLE.

Description of Event or Problem · 1

COUNTY OF COMPLAINT: (B)(6). COMPLAINT STATES: KNOT SECURITY IS DECAYED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
518794 NOVOSYN VIOLET 2/0 (3) 6X45CM SYNTHETIC ABSORBABLE SUTURE GAM B. BRAUN SURGICAL S.A.. C0058646 114223

Patients

Seq Age Sex Outcome Treatment
1 Other