DIMENSION EXL 200
Report
- Report Number
- 1226181-2014-00460
- Event Type
- Malfunction
- Date Received
- September 10, 2014
- Date of Event
- August 4, 2014
- Report Date
- August 15, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JJE
- PMA / PMN Number
- K073604
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE CUSTOMER CONTACTED THE SIEMENS CUSTOMER CARE CENTER (CCC). THE CCC SPECIALIST EVALUATED THE FILTER DATA FOR THE SAMPLE AND DETERMINED THAT NO SAMPLE HAD BEEN ALIQUOTTED FOR THE DILUTE TEST. THE CCC SPECIALIST INFORMED THE CUSTOMER THAT THEY WOULD NOT RECEIVE A NUMERICAL DILUTED RESULT IF THE SAMPLE WAS REMOVED FROM ITS POSITION PRIOR TO DILUTION. THE CUSTOMER WAS ADVISED TO VERIFY THAT OPERATORS WERE TRAINED NOT TO TOUCH ANY ON-BOARD SAMPLES WHILE THE SYSTEM WAS RUNNING. THE CAUSES OF THE INSTRUMENT NOT PRODUCING A NUMERICAL VALUE FOR HCG AND THE OPERATOR REPORTING THE HIL RESULT AS AN HCG RESULT WERE DUE TO USER ERROR. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.
A PATIENT SAMPLE WAS TESTED FOR HUMAN CHORIONIC GONADOTROPIN (HCG) ON A DIMENSION EXL 200 INSTRUMENT. THE SAMPLE RESULTED ABOVE THE DILUTION POINT AND WAS REPEATED UPON DILUTION. THE INSTRUMENT DID NOT PRODUCE A NUMERICAL VALUE FOR THE DILUTED RESULT. THE SAMPLE WAS ALSO TESTED FOR INTERFERENCE DUE TO HEMOLYSIS, ICTERUS, AND LIPEMIA (HIL). THE OPERATOR THEN TOOK THE HIL VALUE AND MANUALLY ENTERED IT INTO THE FIELD FOR THE DILUTED HCG RESULT, WHICH WAS SENT TO THE LABORATORY INFORMATICS SYSTEM (LIS). THE INACCURATE RESULT WAS REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS NOT REPEATED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE HIL RESULT BEING REPORTED TO THE PHYSICIAN(S) AS AN HCG RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 557584 | DIMENSION EXL 200 | CLINICAL CHEMISTRY ANALYZER | JJE | SIEMENS HEALTHCARE DIAGNOSTICS INC. | DIMENSION EXL 200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |