FDA Adverse Event Injury Summary report: N

TENDRIL STS

MDR report key: 4072567 · Received September 9, 2014

Report

Report Number
2017865-2014-16507
Event Type
Injury
Date Received
September 9, 2014
Date of Event
June 30, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVN
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

FINAL ANALYSIS FOUND THAT THE OUTER AND INNER INSULATIONS WERE DAMAGED AT 27.2 CM FROM THE CONNECTOR PIN AS A RESULT OF CLAVICULAR CRUSH. ELECTRICAL TESTS CONFIRMED LOW LEAD IMPEDANCE AND SHORTING AT THIS LOCATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT IN CLINIC, THE RIGHT VENTRICULAR LEAD EXHIBITED LOW LEAD IMPEDANCE AND HIGH CAPTURE THRESHOLDS. UPON EXTRACTION, A LEAD FRACTURE DUE TO CLAVICULAR CRUSH WAS OBSERVED. THE LEAD WAS EXPLANTED AND REPLACED WITHOUT COMPLICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
554722 TENDRIL STS PERMANENT PACEMAKER ELECTRODE NVN ST. JUDE MEDICAL, INC., CRMD 2088TC/58 NA

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention