- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: FINAL ANALYSIS CONFIRMED THAT IT WAS DIFFICULT TO INSERT THE TEST LEAD INTO THE PULSE GENERATORS HEADER AND THE LEAD INSERTION FORCE USED TO INSERT THE LEAD WAS OUT OF SPECIFICATION FOR THE PRODUCT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PULSE GENERATOR CONNECTOR WOULD NOT ACCEPT THE LEAD. THE DEVICE WAS NOT USED AND A NEW DEVICE WAS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 555743 | ASSURITY DR RF | IMPLANTABLE PACEMAKER PULSE GENERATOR | DXY | ST. JUDE MEDICAL, INC., CRMD | PM2240 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR |