FDA Adverse Event Malfunction Summary report: N

ASSURITY DR RF

MDR report key: 4072321 · Received September 9, 2014

Report

Report Number
2017865-2014-16304
Event Type
Malfunction
Date Received
September 9, 2014
Date of Event
May 12, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
DXY
PMA / PMN Number
P880086
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: FINAL ANALYSIS CONFIRMED THAT IT WAS DIFFICULT TO INSERT THE TEST LEAD INTO THE PULSE GENERATORS HEADER AND THE LEAD INSERTION FORCE USED TO INSERT THE LEAD WAS OUT OF SPECIFICATION FOR THE PRODUCT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PULSE GENERATOR CONNECTOR WOULD NOT ACCEPT THE LEAD. THE DEVICE WAS NOT USED AND A NEW DEVICE WAS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
555743 ASSURITY DR RF IMPLANTABLE PACEMAKER PULSE GENERATOR DXY ST. JUDE MEDICAL, INC., CRMD PM2240 NA

Patients

Seq Age Sex Outcome Treatment
1 83 YR